The Ivoclar Programat P90 is a microprocessor-controlled dental ceramic furnace featuring 90 programs, a membrane-sealed keypad for operation, and a 16.5 kg design. It operates via a 7-segment display with automatic head controls, requiring regular calibration via the silver test method for optimal performance. For detailed specifications and operating procedures, see Programat P90 on Bimedis . Programat - Ivoclar Vivadent - PDF Catalogs

3. Key Sections in the Manual You Need to Know

A. Creating Firing Programs

Here is a detailed review of the Ivoclar Programat P90 Manual Press/Furnace. Ivoclar Programat P90 Manual

7. Conclusion

  1. Navigate to Service Menu (Press "Stop" + "Temp" simultaneously).
  2. Select "Offset."
  3. The manual recommends using the Ivoclar Calibration Kit (Ref. 653305AA).
  4. Standard offset: If the furnace reads 850°C but the calibration ceramic shows 860°C, enter +10 in the offset menu.

: The most current official software updates, operating instructions, and safety data sheets are available directly from the Ivoclar Download Center Essential Maintenance & Features Temperature Calibration Navigate to Service Menu (Press "Stop" + "Temp"

The Ivoclar Programat P90 is a microprocessor-controlled dental ceramic furnace featuring 90 programs, a membrane-sealed keypad for operation, and a 16.5 kg design. It operates via a 7-segment display with automatic head controls, requiring regular calibration via the silver test method for optimal performance. For detailed specifications and operating procedures, see Programat P90 on Bimedis . Programat - Ivoclar Vivadent - PDF Catalogs

3. Key Sections in the Manual You Need to Know

A. Creating Firing Programs

Here is a detailed review of the Ivoclar Programat P90 Manual Press/Furnace.

7. Conclusion

  1. Navigate to Service Menu (Press "Stop" + "Temp" simultaneously).
  2. Select "Offset."
  3. The manual recommends using the Ivoclar Calibration Kit (Ref. 653305AA).
  4. Standard offset: If the furnace reads 850°C but the calibration ceramic shows 860°C, enter +10 in the offset menu.

: The most current official software updates, operating instructions, and safety data sheets are available directly from the Ivoclar Download Center Essential Maintenance & Features Temperature Calibration